Factor H Level
B-1H, Radial Immuno Assay, RIA
Test Codes
Order Reference Lab Miscellaneous Send Out: EPIC: LAB848, Beaker: Ref. Lab. Misc., National Jewish Health (NJH): FHL
Department
Send Outs
Specimen Collection Criteria
Collect: One Lavender-top EDTA tube.
Critical frozen. Send specimen for processing immediately after collection.
Physician Office/Draw Specimen Preparation
Critical frozen. Centrifuge to separate plasma from cells within 30 minutes of collection. Transfer plasma to a plastic transport tube and freeze (-20°C/-4°F or below) immediately.
Preparation for Courier Transport
Transport: 1.0 mL plasma, frozen (-20°C/-4°F or below). (Minimum: 0.25 mL)
Rejection Criteria
- Serum specimens.
- Specimens not collected and processed as indicated.
In-Lab Processing
Critical frozen. Centrifuge to separate plasma from cells within 30 minutes of collection. Transfer plasma to a plastic transport tube and freeze (-20°C/-4°F or below) immediately.
Transport: 1.0 mL plasma, frozen (-20°C/-4°F or below). (Minimum: 0.25 mL)
Storage
Specimen Stability for Testing:
Room Temperature (20-26°C or 68-78.8°F): Unacceptable
Refrigerated (2-8°C or 36-46°F): Unacceptable
Frozen (-20°C/-4°F or below): 1 year
Specimen Storage in Department Prior to Disposal:
Specimen retention time is determined by the policy of the reference laboratory. Contact the Send Outs Laboratory with any questions.
Laboratory
Sent to National Jewish Health, Denver, CO.
Performed
Monday – Friday.
Results available in 30 days.
Reference Range
160 - 412 mcg/mL.
Test Methodology
Radial Immunodiffusion (RID)
Interpretation
By report.
CPT Codes
86160